Key result
The AutoCapture algorithm was highly compatible with true bipolar, dual-coil ICD leads, being recommended in 97.8% of patients evaluated.
Why the study?
Does the AutoCapture algorithm demonstrate compatibility with true bipolar, dual-coil ICD leads in patients undergoing ICD implantation?
Observational (n=46)
Yes
Does the AutoCapture algorithm demonstrate compatibility with true bipolar, dual-coil ICD leads in patients undergoing ICD implantation?
The AutoCapture algorithm is highly compatible with true bipolar, dual-coil ICD leads, which may allow for improved generator longevity.
Answers are grounded in this paper and cite their sources.
Supports AutoCapture use with these leads for possible longevity gains; leaves open randomized confirmation of clinical benefit.
AutoCapture (AC) can confirm ventricular capture with true bipolar single coil leads of implantable cardioverter defibrillators (ICD). The compatibility of AC with a new, true bipolar, dual-coil ICD lead needed to be evaluated. This multicenter study enrolled 46 patients (69 +/- 10 years, 37 men) undergoing ICD implantation. All patients received a true bipolar, dual-coil lead. Evoked response (ER) sensitivity and AC threshold tests were performed using a pulse generator with the AC algorithm. Mean capture threshold was 0.85 +/- 0.67 V, pacing impedance 612 +/- 225 Omega, R wave amplitude 13.85 +/- 6.17 mV, and defibrillation threshold 14.4 +/- 5.1 J. AC was recommended in 45 patients (97.8%) with ER and polarization values of 14.86 +/- 7.32 mV and 0.87 +/- 0.69 mV, respectively. The AC algorithm was highly compatible with true bipolar, dual-coil ICD leads. An AC algorithm specifically designed for an ICD may improve the generator longevity. Further examination of AC compatibility with other leads is warranted.
No takes yet. Share an insight, caveat, or question.
Acosta et al. (2005) conducted an observational in Patients undergoing ICD implantation (n=46). AutoCapture algorithm with true bipolar, dual-coil ICD lead was evaluated on AutoCapture recommendation based on evoked response sensitivity and threshold tests. The AutoCapture algorithm was highly compatible with true bipolar, dual-coil ICD leads, being recommended in 97.8% of patients evaluated.