Key result
The bipolar ventricular pacing lead IsoFlex S 1636/1646T showed good electrical performance over 12 months, with AutoCapture activation recommended in at least 95% of patients.
Why the study?
Does the bipolar ventricular pacing lead IsoFlex S demonstrate adequate electrical performance and compatibility with the AutoCapture algorithm?
Observational (n=252)
Yes
Does the bipolar ventricular pacing lead IsoFlex S demonstrate adequate electrical performance and compatibility with the AutoCapture algorithm?
The IsoFlex S bipolar ventricular pacing lead demonstrates stable electrical performance and high compatibility with the AutoCapture algorithm over 1 year.
Answers are grounded in this paper and cite their sources.
May support IsoFlex S lead selection in pacing; leaves open randomized comparisons of long-term performance.
BACKGROUND: Bipolar low polarization electrodes are recommended for a regular AutoCapture (St. Jude Medical, Inc., Sylmar, CA, USA) function in order to effectively detect the evoked response (ER) signal. The objective of this national multicenter registry was to evaluate the electrical performance and the AutoCapture characteristics of the bipolar ventricular pacing lead IsoFlex S, model 1636T or 1646T (St. Jude Medical), in combination with single- and dual-chamber pacemakers. METHODS: Ventricular pacing and sensing thresholds, lead impedance, ER amplitude, and polarization signals were measured at discharge and routine follow-up visits after 1, 3, 6, 9, and 12 months. AutoCapture activation was recommended based on the results of the ER sensitivity test. RESULTS: Of the 252 patients initially included, 109 (43%) have completed the follow-up. The mean ventricular pacing threshold was 0.43 +/- 0.19 V at discharge and 0.68 +/- 0.32 V at 12 months postimplant. The values for the ventricular sensing threshold were between 9.51 +/- 4.12 and 9.99 +/- 4.09 mV at discharge and at the 12-month follow-up. The unipolar lead impedance decreased from 533 +/- 94 to 476 +/- 73 ohms during the follow-up. The mean ER amplitude was 16.47 +/- 6.70 mV at discharge and 17.42 +/- 7.43 mV after 12 months, and the corresponding mean polarization signals were 0.59 +/- 1.00 and 0.74 +/- 1.24 mV, respectively. AutoCapture activation was recommended in at least 95% of the patients investigated over the 12-month follow-up. CONCLUSION: The bipolar ventricular pacing lead IsoFlex S 1636/1646T shows a good electrical performance and is mostly compatible with the AutoCapture algorithm.
No takes yet. Share an insight, caveat, or question.
Lotze et al. (2009) conducted an observational in Patients with single- and dual-chamber pacemakers (n=252). Bipolar ventricular pacing lead IsoFlex S (model 1636T or 1646T) was evaluated on Electrical performance and AutoCapture characteristics. The bipolar ventricular pacing lead IsoFlex S 1636/1646T showed good electrical performance over 12 months, with AutoCapture activation recommended in at least 95% of patients.
Synapse has enriched 10 related papers. Consider them for comparative context: