Objective: To assess the accuracy of a skin-mounted device to measure lumbar spinal rotations in rowers. Methods: Nine subjects were each imaged in 12 extremes of lateral flexion and rotation in an open MR scanner, and the movement between external MR markers fixed to the skin, and the underlying vertebral bodies was quantified. Results: The average error in measuring the changes in the lateral flexion and rotation between the MR markers and the underlying vertebral bodies was ±5.0° (95% CI 3.5–6.4°, S.E. 0.7°) for lateral flexion and 4.4° (95% CI 3.2–5.6°, S.E 0.6°) for lateral rotation (nine subjects). Conclusions: This study has demonstrated the accuracy of using an externally mounted device to measure out-of-plane rotations for the extremes of motion achieved during a rowing activity, allowing this system to be used to quantify deviations in rowing technique. The method could be applied to other repetitive activities of the spine.