High impedance leads are better described as high current density and high efficiency leads, and small tip electrodes should be combined with low polarization surface treatment for safety and energy saving.
Answers are grounded in this paper and cite their sources.
May support refined safety margin programming for energy efficiency in pacing; leaves open clinical outcomes with high-current-density leads.
For physicians who are not familiar with the electrical basis of cardiac stimulation, high impedance leads previously (10 years ago) considered as bad electrodes (high voltage) are now considered as the 'epitome' of lead technology (low energy drain): clearly impedance is not a good parameter for characterizing the qualities of a pacing lead. Using a simplified approach to the electrostimulation 'paradigm', it is easy to establish that modern high impedance leads are in fact high current density leads and high efficiency leads (better description). It is also possible to establish that routine programming of the safety margin at 100% above threshold parameters is associated with a decrease in the penetration of the electric field according to the reduction of the cathode surface area. For safety and energy saving, a small tip electrode could be combined with a low polarization surface treatment and a reduction in fibrosis development between electrode and myocardium.