Does a three-monthly laboratory test detect impending failure in implanted pacemakers to allow elective replacement?
Observational (n=19)
Does a three-monthly laboratory test detect impending failure in implanted pacemakers to allow elective replacement?
Answers are grounded in this paper and cite their sources.
A simple laboratory test performed every three months can reliably detect impending pacemaker failure, allowing for elective rather than emergency replacement.
As implantable pacemakers last a variable time, it is desirable to be able to foretell when each individual pacemaker is nearing the end of its life. A test which detects a number of different impending electronic failures including battery depletion is described. The test requires only standard laboratory equipment and has been used for four years as a three-monthly procedure in an out-patient clinic. In 17 of 19 consecutive patients the test has enabled a change of pacemaker to be carried out electively. rather than after an arbitrary period or as an emergency when pacing has failed.
No takes yet. Share an insight, caveat, or question.
Davies et al. (1969) conducted an observational in Implanted pacemakers (n=19). Test to detect impending electronic failures including battery depletion was evaluated on Elective change of pacemaker before failure.