Why the study?
Does regular electronic testing predict generator failure in patients with implanted cardiac pacemakers?
Does regular electronic testing predict generator failure in patients with implanted cardiac pacemakers?
Regular electronic testing of implanted pacemakers can predict battery-related generator failure, allowing for optimal resource use and minimizing unexpected failures.
Answers are grounded in this paper and cite their sources.
Stimulus artefact analysis may predict battery-related pacemaker generator failure; leaves open whether routine testing prevents unexpected failures in practice.
Seventy patients with implanted cardiac pacemakers have been followed in a special pacemaker clinic for periods up to five years. A system of electronic testing has been used after implanting 142 generators. Analysis of the stimulus artefact made it possible to predict generator failure in the majority of systems. In those pacemakers where the failure could be predicted, the fault was always associated with the battery supply. It was not possible to predict sudden electronic component failure. In order to use the generator's full resources and minimize unexpected battery failure, regular electronic testing is recommended.
No takes yet. Share an insight, caveat, or question.
Smith et al. (1971) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: