Why the study?
Does routine pacemaker control with selective replacement improve pulse generator lifetime in patients with permanent pacemakers?
Does routine pacemaker control with selective replacement improve pulse generator lifetime in patients with permanent pacemakers?
Routine pacemaker control with selective replacement extends pulse generator lifetime by an average of 6.5 months compared to elective replacement at 24 months.
May support selective replacement policies in pacemakers; leaves open whether routine control improves outcomes without randomized confirmation.
Two hundred and fifty patients with permanent pacemaker have been followed up with routine pacemaker controls in a pacemaker clinic for a 21 months period 1/3-73 to 1/1-75. Ninety-five pulsegenerators were replaced. Sixty-two of the replacements were due to impending battery exhaustion, nine elective and 24 for other reasons. Signs discovered by the patients led to replacements in 21 cases for impending battery exhaustion and in 17 cases for other reasons. Forty-one replacements for impending battery exhaustion and 7 other replacements followed a scheduled visit to the pacemaker clinic. The selective replacement policy resulted in an average gain of pulsegenerator lifetime of 6.5 months, compared to a 24 minths elective replacement policy. Increased safety is obtained by routine control of pacemaker patients in a pacemaker clinic.
No takes yet. Share an insight, caveat, or question.
H Grendahl (1976) studied this question.