Routine pacemaker follow-up detects system malfunctions in ~6% of patients, predominantly correctable sensing issues.
Why the study?
Specialized pacemaker follow-up is necessary to detect and correct all malfunctions of the pacing system beyond battery depletion, despite improved pulse generator reliability and longevity.
Observational (n=1,065)
Malfunction rates vary by implant duration; supports vigilant follow-up but leaves open optimal surveillance strategies in modern devices.
The goal of pacemaker follow-up is not only to detect battery depletion but also to detect all malfunctions of the pacing system and, when possible, to correct such problems using programming. During one year, we discovered 61 such malfunctions in a clinic of 1065 patients (5.7%). These were more frequent in the first year (7.7%) than in the third to fifth years of follow-up (range 3.1-4.8%). The incidence rose again in the sixth and subsequent years (7-7.7%). Despite a significant occurrence of malfunctions (5.2%) among multiprogrammable pacemakers, the necessity for operative intervention for their correction was low (1.2%). Sensing problems were the most common (57%) and the most likely to be corrected by reprogramming (85%); problems involving loss of capture were less likely to be corrected by programming (38.5%). Battery depletion accounted for only 18% of malfunctions, occurring earliest in the forty-third month of follow-up. Pulse generator longevity of those devices reaching end of battery life during the study period was 68.6 +/- 16.7 months (mean + SD). We conclude that specialized pacemaker follow-up continues to be necessary despite improved pulse generator reliability and longevity. Indeed, with reprogramming, it presently plays an even more important role than in the past. Follow-up should be oriented not only to the detection of battery depletion but also toward a comprehensive surveillance of pacemaker system function.
No takes yet. Share an insight, caveat, or question.
Griffin et al. (1986) conducted an observational in Pacemaker patients (n=1,065). Pacemaker follow-up was evaluated on Pacemaker system malfunctions. Routine pacemaker follow-up over one year detected system malfunctions in 5.7% of patients, with sensing problems being the most frequent (57%) and highly correctable by reprogramming (85%).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: