Key result
The HemosIL AcuStar D-Dimer immunoassay demonstrated remarkable analytical accuracy against Vidas D-dimer (AUC 0.998, P<0.001) with optimal sensitivity and specificity at a 466 ng/ml threshold.
Why the study?
Does the HemosIL AcuStar D-Dimer immunoassay demonstrate acceptable analytical performance compared to established D-dimer assays?
Observational
Does the HemosIL AcuStar D-Dimer immunoassay demonstrate acceptable analytical performance compared to established D-dimer assays?
Effect estimate: AUC 0.998
p-value: p=<0.001
The novel HemosIL AcuStar D-Dimer immunoassay demonstrates high accuracy, precision, and excellent agreement with reference assays, supporting its use in clinical laboratories.
May support lab adoption of HemosIL AcuStar D-dimer; leaves open clinical outcome validation in patients.
Several lines of clinical evidence as well as guidelines and recommendations suggest that the overall diagnostic performance of D-dimer testing outstrips that of any other biomarker in the diagnostic approach of patients with venous thromboembolism or disseminated intravascular coagulation. Along with specific technical characteristics, the analytical performance of each D-dimer immunoassay should, however, be assessed before implementation in clinical practice. The aim of this study was to evaluate the analytical performance of HemosIL AcuStar D-Dimer immunoassay, a novel chemiluminescent immunoassay specifically designed for the instrument Instrumentation Laboratory ACL AcuStar. The within and between-run imprecision (n = 20) was comprised between 3.6 and 5.8%. The linearity was excellent up to 16 200 ng/ml (r = 1.00; P < 0.001), and optimal between 224 and 69 900 ng/ml (r = 0.992; P < 0.001). A significant agreement of values was observed between HemosIL AcuStar D-Dimer and HemosIL D-Dimer HS for ACL TOP (r = 0.884; P < 0.001), as well as with Vidas D-Dimer (r = 0.791; P < 0.001). Results of HemosIL AcuStar D-Dimer displayed a modest negative bias as compared with those of Vidas D-dimer (mean bias -22%; 95% Confidence Interval, -122% to 49%). The analytical accuracy assessed against Vidas D-dimer also yielded an area under the curve of 0.998 (P < 0.001). As regards, the preliminary definition of cutoff value, optimal sensitivity (100%) and specificity (96%) were found at a diagnostic threshold of 466 ng/ml. The results of this investigation attest that the novel HemosIL AcuStar D-Dimer is characterized by remarkable accuracy and precision, optimal linearity and excellent agreement with the reference immunoassay. As such, its technical and analytical performance would make it a suitable method for the rapid and accurate quantification of D-dimer in clinical laboratories.
No takes yet. Share an insight, caveat, or question.
Lippi et al. (2012) reported an observational. HemosIL AcuStar D-Dimer immunoassay vs. Vidas D-dimer and HemosIL D-Dimer HS was evaluated on Analytical accuracy assessed against Vidas D-dimer (Area Under the Curve) (AUC 0.998, p=<0.001). The HemosIL AcuStar D-Dimer immunoassay demonstrated remarkable analytical accuracy against Vidas D-dimer (AUC 0.998, P<0.001) with optimal sensitivity and specificity at a 466 ng/ml threshold.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: