Key result
In patients receiving the CENTERA transcatheter heart valve, a perimeter oversizing cut-off of 15.9% predicted ≥mild paravalvular aortic regurgitation.
Why the study?
Does the degree of CT-based annular perimeter oversizing affect the incidence of paravalvular aortic regurgitation and permanent pacemaker implantation in patients receiving the CENTERA transcatheter heart valve?
Observational (n=198)
Yes
Does the degree of CT-based annular perimeter oversizing affect the incidence of paravalvular aortic regurgitation and permanent pacemaker implantation in patients receiving the CENTERA transcatheter heart valve?
The CENTERA transcatheter heart valve demonstrates a wide sizing tolerance, with an optimal perimeter oversizing cut-off of 15.9% to mitigate paravalvular regurgitation without increasing pacemaker risk.
Answers are grounded in this paper and cite their sources.
May inform oversizing strategies to lower paravalvular leak risk with CENTERA valves; leaves open prospective validation.
AIMS: We aimed to evaluate the relationship between CT-based annular perimeter oversizing and the incidence of paravalvular aortic regurgitation (PAR) and permanent pacemaker (PPM) implantation in patients treated with the new self-expanding CENTERA transcatheter heart valve (THV) for severe aortic stenosis. METHODS AND RESULTS: One hundred and ninety-eight patients in the CENTERA-EU trial were stratified a priori into four groups based on the perimeter oversizing (2.5-10%, 10-15%, 15-20% and >20%). PAR at 30 days was moderate or higher in 0.6% of patients. The frequency of PPM implantation was 4.9%. The mean perimeter oversizing was 16.2±5.6%. For patients with a perimeter oversizing >10%, an inverse relationship between oversizing and ≥mild PAR was observed (43.3% for 10-15% oversizing; 37.7%, 15-20%; 33.3%, >20%). No association between oversizing and effective orifice area was observed. The optimal cut-off value of perimeter oversizing for the prediction of ≥mild PAR was 15.9% (AUC 0.718, 95% CI: 0.576, 0.860). No annular ruptures were observed. CONCLUSIONS: The CENTERA THV appears to have a wide range of sizing tolerance. The degree of oversizing to mitigate PAR is relatively low compared to other self-expanding transcatheter devices. There appears to be no compromise between occurrence of PAR and PPM across this wide range of oversizing.
No takes yet. Share an insight, caveat, or question.
Kim et al. (2018) conducted an observational in Severe aortic stenosis (n=198). CT-based annular perimeter oversizing vs. Different degrees of oversizing (2.5-10%, 10-15%, 15-20% and >20%) was evaluated on Prediction of ≥mild paravalvular aortic regurgitation (PAR) (AUC 0.718, 95% CI 0.576-0.860). In patients receiving the CENTERA transcatheter heart valve, a perimeter oversizing cut-off of 15.9% predicted ≥mild paravalvular aortic regurgitation.