The relaxation phenomena developing after the injection of the sample into the channel during Field-Flow Fractionation were analyzed theoretically. Relaxation times, necessary to stop the flow immediately after the injection, were calculated on the basis of various models describing the relaxation processes. A stop-flow time long enough so that the deviation from the equilibrium would not influence the retention within the limits of experimental errors and so that the maximum efficiency would be reached was required. Sedimentation Field-Flow Fractionation was used to verify the theoretical analysis. The diffusion coefficient of latex used for the study is small and thus the long relaxation times allow to reach sufficient precision of the measurement. The experiments performed at different flow rates and using different ways of sampling proved a good agreement between the experimental relaxation times and the values calculated theoretically. The stop-flow technique was proved both theoretically and experimentally to be advantageous from the viewpoint of shortening the time of fractionation The disagreement with the previous experimental data obtained by other authors was explained. There was a good agreement between our experimental data and the theoretical predictions. Field-Flow Fractionation using the stop-flow technique at the constant field operation was found to be very suitable for exact physico-chemical measurements.