Key result
A soft-tip ablation catheter became entrapped in the mitral valve apparatus during a remote magnetic navigation ablation procedure, requiring surgical removal.
Case Report (n=1)
This case highlights the need for closer fluoroscopic monitoring when performing catheter loops inside the ventricles using the RMN system to prevent catheter entrapment in the mitral valve apparatus.
Answers are grounded in this paper and cite their sources.
First reported mitral valve entrapment during RMN ablation urges fluoroscopic caution; leaves open true incidence and prevention strategies.
Remote magnetic navigation (RMN) is as safe and effective as manual navigation for catheter ablation of ventricular arrhythmias. This case is the first description of a soft-tip ablation catheter entrapment in the mitral valve apparatus during an RMN ablation procedure. The tight knot created by the catheter around a mitral valve chordae required surgical removal. This complication, which has never been reported before, highlights the need for closer fluoroscopic monitoring when performing catheter loops inside the ventricles when using the RMN system.
No takes yet. Share an insight, caveat, or question.
Bénali et al. (2021) conducted a case report in Ventricular arrhythmias (n=1). Remote magnetic navigation (RMN) ablation procedure was evaluated on Catheter entrapment in the mitral valve apparatus. A soft-tip ablation catheter became entrapped in the mitral valve apparatus during a remote magnetic navigation ablation procedure, requiring surgical removal.