Key result
Intense cardiac electrical stimulation (6-8 V) via a pacemaker or CRT device reproducibly induced cardiac sensation in 6 of 15 patients and elicited measurable cerebral evoked potentials in all patients.
Why the study?
Does intense cardiac electrical stimulation provoke brain responses in patients with implanted pacemakers or CRT devices?
Cross-Sectional (n=15)
No
Does intense cardiac electrical stimulation provoke brain responses in patients with implanted pacemakers or CRT devices?
This study provides the first evidence of brain responses to cardiac electrical stimulation in humans, offering a novel EEG-based method to evaluate cardiac sensation.
May enable EEG-based assessment of device-related cardiac sensation; hypothesis-generating and should not yet change practice.
Although cardiac sensation, such as palpitation or chest pain, is common and is sometimes a malignant sign of heart diseases, the mechanism by which the human brain responds to afferent signals from the heart remains unclear. In this study, we investigated whether electrical stimulation of the heart provokes brain responses in humans. We examined 15 patients (age: 65.4 ± 3.1 years old, 11 males and 4 females) implanted with either a cardiac pacemaker or cardiac resynchronization therapy (CRT) device. Electroencephalogram (EEG) was simultaneously recorded from the vertex during right ventricular pacing at 70-100 beats/min at baseline (1.5 V) and intense (6-8 V) stimulation sessions. We evaluated brain responses to cardiac electrical stimulation by measuring cerebral potentials that were obtained by subtracting the average of 100 EEG waves triggered by cardiac pacing during baseline stimulation from those during the intense stimulation. Intense stimulation of the cardiac pacemaker or CRT device reproducibly induced cardiac sensation in 6 out of the 15 patients; namely, the remaining 9 patients showed no reproducible response. Brain responses were evident by averaging cerebral potentials from all of the 15 patients and those from 6 patients with reproducible cardiac sensation. To the best our knowledge, this is the first report that demonstrates the brain responses to cardiac electrical stimulation in humans. This new method should be useful for examining pathophysiology of cardiac diseases with pathological cardiac sensation, including cardiac anxiety and silent myocardial ischemia.
No takes yet. Share an insight, caveat, or question.
Suzuki et al. (2011) conducted a cross-sectional in Patients with implanted cardiac pacemaker or CRT device (n=15). Intense cardiac electrical stimulation vs. Baseline stimulation (1.5 V) was evaluated on Cerebral evoked potentials (CEPs) and cardiac sensation. Intense cardiac electrical stimulation (6-8 V) via a pacemaker or CRT device reproducibly induced cardiac sensation in 6 of 15 patients and elicited measurable cerebral evoked potentials in all patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: