Key result
The Accutracker II (version 30/23) ambulatory blood pressure monitor fulfilled AAMI criteria and was graded A for systolic and C for diastolic blood pressure by the BHS system.
Why the study?
Does the Accutracker II (version 30/23) ambulatory blood pressure monitor accurately measure blood pressure compared to observer measurements according to AAMI and BHS protocols?
Population
120 subjects
Comparison
Accutracker II ambulatory blood pressure monitor vs Observer blood pressure measurements
Design
Other
Follow-up
24 hours (for field evaluation)
Authors
Loading...
Accutracker II meets AAMI/BHS validation thresholds; leaves open questions on real-world performance and protocol differences.
Observational (n=120)
Does the Accutracker II (version 30/23) ambulatory blood pressure monitor accurately measure blood pressure compared to observer measurements according to AAMI and BHS protocols?
Effect estimate: mean difference -2.2 +/- 0.8 / -3.5 +/- 0.9 mmHg
The Accutracker II (version 30/23) ambulatory blood pressure monitor meets AAMI criteria and achieves BHS grade A for systolic and C for diastolic blood pressure, demonstrating acceptable accuracy while highlighting differences between validation protocols.
Taylor et al. (1993) conducted an observational in Blood pressure measurement (n=120). Accutracker II (version 30/23) ambulatory blood pressure monitor vs. Observer measurements was evaluated on Classification of device accuracy according to the criteria of the BHS and AAMI protocols (mean difference -2.2 +/- 0.8 / -3.5 +/- 0.9 mmHg). The Accutracker II (version 30/23) ambulatory blood pressure monitor fulfilled AAMI criteria and was graded A for systolic and C for diastolic blood pressure by the BHS system.