Key result
Current methods for monitoring antiplatelet medication, such as the PFA-100 and standard TEG, lack specificity or sensitivity, highlighting the need for more robust platelet function assessments.
Answers are grounded in this paper and cite their sources.
Highlights the limitations of current platelet function monitoring devices (PFA-100, TEG) for assessing the effects of antiplatelet medication in the perioperative setting.
We read with interest the recent review about thienopyridine antiplatelet drugs. A major problem faced by anaesthetists is the monitoring of the effects of antiplatelet medication. Although the platelet function analyser (PFA-100, Dade Behring, IL, USA) appears to be the answer to this problem, it is a global test of primary haemostasis and is not specific for antiplatelet drugs [1]. The PFA-100 has been used to monitor the antiplatelet effect of aspirin [2]; however, the results may be affected by the citrate concentration in the sampling tube [3]. There is less experience with the use of the PFA-100 to assess the effect of clopidogrel, and it is uncertain whether the citrate concentration is also important in this situation. Results from the PFA-100 may also be affected by smoking, haematocrit and gender [1]. The Thrombelastograph® (TEG®) Haemostasis Analyser (Haemoscope Corp., USA) is a near-patient method of monitoring whole blood coagulation. Standard use of the TEG® system cannot detect the effect of aspirin [4]. However, a new TEG® assay has been developed specifically for this purpose (Personal Communication: Haemoscope Corp., USA). The use of antiplatelet medication has implications for the safety of regional anaesthesia and propensity to bleed during surgery. Further studies are needed to provide a sufficiently robust method of platelet function assessment to guide anaesthetic management.
No takes yet. Share an insight, caveat, or question.
R. Self (2003) conducted a letter in Antiplatelet medication monitoring. Platelet function assessment (PFA-100, TEG) was evaluated. Current methods for monitoring antiplatelet medication, such as the PFA-100 and standard TEG, lack specificity or sensitivity, highlighting the need for more robust platelet function assessments.
Synapse has enriched 10 related papers. Consider them for comparative context: