This is the authors' abstract. We don't add key points for this paper.
Enhancement of X-ray excited optical luminescence in a 100 µm-thick diamond plate by introduction of defect states via electron beam irradiation and subsequent high-temperature annealing is demonstrated. The resulting X-ray transmission-mode scintillator features a linear response to incident photon flux in the range 7.6 × 108 to 1.26 × 1012 photons s−1 mm−2 for hard X-rays (15.9 keV) using exposure times from 0.01 to 5 s. These characteristics enable a real-time transmission-mode imaging of X-ray photon flux density without disruption of X-ray instrument operation.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Stoupin et al. (2020) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: