Key result
The MULTIPLE optical platform enabled accurate measurement of action potential dynamics across a 24-well plate with interwell variability of approximately 10%.
Why the study?
Current optical techniques for fast characterization of cardiac electrophysiology rely on serial well-by-well automated fluorescence microscopy, prompting the need for an alternative cost-effective approach for parallel detection across multiwell plates.
The MULTIPLE optical system provides a cost-effective, parallel detection method for high-throughput in vitro cardiac electrophysiological investigations.
May advance high-throughput in vitro cardiac electrophysiology; leaves open translation to human models.
Current techniques for fast characterization of cardiac electrophysiology employ optical technologies to control and monitor action potential features of single cells or cellular monolayers placed in multiwell plates. High-speed investigation capacities are commonly achieved by serially analyzing well after well employing fully automated fluorescence microscopes. Here, we describe an alternative cost-effective optical approach (MULTIPLE) that exploits high-power LED arrays to globally illuminate a culture plate and an sCMOS sensor for parallel detection of the fluorescence coming from multiple wells. MULTIPLE combines optical detection of action potentials using a red-shifted voltage-sensitive fluorescent dye (di-4-ANBDQPQ) with optical stimulation, employing optogenetic actuators, to ensure excitation of cardiomyocytes at constant rates. MULTIPLE was first characterized in terms of interwell uniformity of the illumination intensity and optical detection performance. Then, it was applied for probing action potential features in HL-1 cells (i.e., mouse atrial myocyte-like cells) stably expressing the blue light-activatable cation channel CheRiff. Under proper stimulation conditions, we were able to accurately measure action potential dynamics across a 24-well plate with variability across the whole plate of the order of 10%. The capability of MULTIPLE to detect action potential changes across a 24-well plate was demonstrated employing the selective K v 11.1 channel blocker (E-4031), in a dose titration experiment. Finally, action potential recordings were performed in spontaneous beating human induced pluripotent stem cell derived cardiomyocytes following pharmacological manipulation of their beating frequency. We believe that the simplicity of the presented optical scheme represents a valid complement to sophisticated and expensive state-of-the-art optical systems for high-throughput cardiac electrophysiological investigations.
No takes yet. Share an insight, caveat, or question.
Credi et al. (2021) studied Cardiac electrophysiology (in vitro). MULTIPLE optical platform was evaluated on Action potential dynamics and interwell uniformity. The MULTIPLE optical platform enabled accurate measurement of action potential dynamics across a 24-well plate with interwell variability of approximately 10%.