Key result
Lead shielding was accurately modeled and demonstrated to be an effective and practical material for reducing fetal dose during CT pulmonary angiography.
Why the study?
Does lead shielding effectively reduce fetal radiation dose during CT pulmonary angiography in a phantom model?
Does lead shielding effectively reduce fetal radiation dose during CT pulmonary angiography in a phantom model?
Conventional lead aprons provide significant and effective shielding for the fetus during CT pulmonary angiography, though dedicated shields could further optimize protection and comfort.
Answers are grounded in this paper and cite their sources.
Supports lead shielding for fetal dose reduction in CTPA models; leaves open clinical validation in pregnancy.
The aim of this work is to construct and validate a model to describe the variation in fetal dose as a function of the thickness of abdominal lead shielding used during CT pulmonary angiography and to determine the optimal shielding material. An anthropomorphic phantom was modified to contain a 15 cm(3) ionization chamber at the site of the uterus. Fetal dose was measured with varying thicknesses of lead shielding at four values of tube potential (kV(p)). Data generated by the proposed model were compared with experimental data to determine the validity of the model. The effect of lead shielding has been modelled accurately and results have shown that, although alternative materials could be used, lead is an effective and practical shielding material. In conclusion, lead remains a suitable shielding material and a pair of conventional lead aprons provides significant shielding for the fetus; we recommend that aprons should be reserved specifically for this purpose. However, it is possible that a dedicated and specifically designed lead shield could reduce fetal dose more effectively whilst also reducing patient discomfort.
No takes yet. Share an insight, caveat, or question.
Iball et al. (2008) studied CT pulmonary angiography during pregnancy. Lead shielding vs. Alternative materials was evaluated on Fetal dose variation as a function of shielding thickness. Lead shielding was accurately modeled and demonstrated to be an effective and practical material for reducing fetal dose during CT pulmonary angiography.
Synapse has enriched 10 related papers. Consider them for comparative context: