Why the study?
What are the complication rates and long-term safety of dual chamber pacing system implantation?
What are the complication rates and long-term safety of dual chamber pacing system implantation?
Improvements in lead design, programmability, and surgical experience have reduced complication rates of dual chamber pacing to acceptable levels, making it as safe as single chamber pacing.
Answers are grounded in this paper and cite their sources.
These rates suggest dual-chamber pacing is feasible with improving safety; leaves open multicenter validation before broader adoption.
In a six-year period (1980-1985), 345 dual chamber pacing systems were implanted in our hospital. Intraoperative complications were rare (n = 5/345 = 1.4%). 25 patients (7%) have died, 11 (3%) were lost to follow-up. A total of 39 patients (13%) showed minor problems, which could be overcome by reprogramming. Another 48 complications (15%) required reoperation. Nearly 70% of those reoperations (n = 31/48) were caused by a malfunction of the atrial lead. This high complication rate of 9% related to all implanted atrial leads could be reduced in the last three years to 2%. We could maintain a dual chamber stimulation mode (DDD, DDI, DVI, VDD) in 396 patients (99%). We conclude that improvements in lead design, pulse generator's programmability, and surgical experience have reduced complication rates to acceptable low values. Meanwhile, dual chamber pacing can be considered as safe as single chamber pacemaker therapy.
No takes yet. Share an insight, caveat, or question.
Markewitz et al. (1986) studied this question.