Key result
A simple and sensitive HPLC-UV method was developed for determining paclitaxel in human and rat blood, with detection limits of 10, 9.5, and 7.5 ng/mL for human plasma, serum, and rat plasma.
A simple and sensitive HPLC-UV method was successfully developed for determining low-dose paclitaxel in human and rat blood samples.
Answers are grounded in this paper and cite their sources.
May enable low-dose paclitaxel monitoring in plasma/serum; leaves open clinical validation before routine use.
A simple and sensitive HPLC-UV method was developed for the determination of paclitaxel (TXL) in human and rat blood samples. 4-Hydroxybenzoic acid n-hexyl ester was used as an internal standard. TXL was extracted by a liquid-liquid extraction with tert-butylmethyl ether. The disturbing peaks in the case of serum sample were removed by pre-extraction with hexane. The separation of TXL was achieved within 25 min using an ODS column with 50% acetonitrile aqueous solution as a mobile phase at a flow rate of 1.0 mL/min. The eluent was monitored at 230 nm, and the resulted retention times of TXL and IS were 11.2 and 20.4 min. The detection limits of TXL for human plasma, serum and rat plasma samples at a signal-to-noise ratio of 3 were 10, 9.5 and 7.5 ng/mL, respectively. The proposed methods were applicable to the determination of TXL in human patients' plasma ranging from 15 to 27 ng/mL. Furthermore, monitoring of the time course of TXL after its single administration to rat could be demonstrated.
No takes yet. Share an insight, caveat, or question.
Yonemoto et al. (2007) studied this question. HPLC-UV method for paclitaxel determination was evaluated on Detection limits of paclitaxel. A simple and sensitive HPLC-UV method was developed for determining paclitaxel in human and rat blood, with detection limits of 10, 9.5, and 7.5 ng/mL for human plasma, serum, and rat plasma.
Synapse has enriched 10 related papers. Consider them for comparative context: