Key result
A laptop computer placed on the chest induced an inappropriate magnet-mode switch in an implantable cardioverter-defibrillator due to the static magnetic field from its hard disk.
Case Report (n=1)
Placing a laptop computer on the chest can cause an ICD to inappropriately switch to magnet mode due to the hard disk's static magnetic field, disabling tachyarrhythmia therapies.
Answers are grounded in this paper and cite their sources.
May warrant caution with laptop proximity to ICDs; case report leaves open the prevalence of such interference.
An implantable cardioverter-defibrillator (ICD) experienced electromagnetic interference from a laptop computer's hard disk. The patient with the ICD was using his laptop computer at home while lying on his bed. The laptop was positioned on his chest, when he heard a beeping sound from the ICD, indicating magnet mode conversion. This situation was replicated in a controlled environment, and the conversion was found to be due to the static magnetic field produced by the laptop's hard disk. The ICD's conversion to magnet mode can be dangerous because it ends all tachyarrhythmia detections and therapies.
No takes yet. Share an insight, caveat, or question.
Tiikkaja et al. (2012) conducted a case report in Implantable cardioverter-defibrillator (n=1). Laptop computer (static magnetic field from hard disk) was evaluated on Magnet mode conversion (inappropriate switch). A laptop computer placed on the chest induced an inappropriate magnet-mode switch in an implantable cardioverter-defibrillator due to the static magnetic field from its hard disk.