Answers are grounded in this paper and cite their sources.
Bis(2-methylbenzaldehyde)butane-2,3-dihydrazone(TDSB) was used as new N-N Schiff's base which plays the role of an excellent ion carrier in the construction of a La(III) membrane sensor. The best performance was obtained with a membrane containing, 30% poly(vinyl chloride), 60% benzyl acetate, 6% TDSB and 4% sodium tetraphenyl borate. This sensor reveals a very good selectivity towards La(III) ions over a wide variety of cations, including alkali, alkaline earth, transition and heavy metal ions. The proposed electrode exhibits a Nernstian behavior (with slope of 19.8 mV per decade) over a wide concentration range (1.0 × 10 ⁻⁵ -1.0 × 10 ⁻¹ M). The detection limit of the sensor is 7.0 × 10 ⁻⁶ M. It has a very short response time, in the whole concentration range ( ~ 5 s), and can be used for at least twelve weeks in the pH range of 3.0-9.4. The proposed sensor was successfully applied as an indicator electrode for the potentiometric titration of a La(III) solution, with EDTA. It was also successfully applied in the determination of fluoride ions in three mouth wash preparations.
No takes yet. Share an insight, caveat, or question.
Mohammad Reza Ganjali (2005) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: