Why the study?
Micra AV provides atrioventricular synchrony, and first-degree AV block can cause a reduction of myocardial contractility.
Does single-lead atrial sensing ventricular pacing (VDD) improve left ventricle global longitudinal strain compared to spontaneous rhythm in patients with first-degree atrioventricular block?
Does single-lead atrial sensing ventricular pacing (VDD) improve left ventricle global longitudinal strain compared to spontaneous rhythm in patients with first-degree atrioventricular block?
In a small cohort of patients with first-degree AV block, leadless VDD pacing significantly improved left ventricular global longitudinal strain compared to spontaneous rhythm.
Answers are grounded in this paper and cite their sources.
VDD pacing was associated with better LV GLS in this case report; hypothesis-generating and should not yet change practice in first-degree AV block.
Aim: Micra AV represents a leadless endocardial pacing system able to detect atrial contractions providing atrioventricular synchrony. A reduction of myocardial contractility may be detected in case of first-degree atrioventricular block (AVB). Materials & methods: In six patients with first-degree AVB (PQ interval ≥220 msec) was evaluated the left ventricle global longitudinal strain (LV GLS) by speckle tracking (ST) echocardiography during single-lead atrial sensing ventricular pacing (VDD) stimulation as compared with spontaneous rhythm (SR), 24–48 h after Micra AV implantation. Results: A statistically significant difference between the two modalities was observed (LV GLS during SR: -14.7% [interquartile range (IQR) 5.5], LV GLS during VDD pacing: -16.1% [IQR 5.2]; p value = 0.041). Conclusion: Our preliminary results suggest an improvement of myocardial contractility with VDD pacing as compared with SR.
No takes yet. Share an insight, caveat, or question.
Guaricci et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: