Key result
Micra AV VDD pacing narrows QRS to ~90 ms in RBBB but widens complexes in LBBB.
Case Report (n=1)
Right ventricular septal pacing with a Micra AV leadless pacemaker may inadvertently correct right bundle branch block (narrowing QRS to 90 ms) but not left bundle branch block (QRS 140 ms), highlighting the importance of pacing site selection.
Answers are grounded in this paper and cite their sources.
Pacing site selection remains critical with leadless devices; leaves open optimal positions for left bundle branch block correction.
A Micra AV leadless pacemaker was implanted via the femoral approach due to superior vena cava occlusion. Follow-up included 12‑lead ECG, device interrogation, and imaging. Electrocardiography revealed alternating bundle-branch block patterns. VDD pacing during right bundle branch block (RBBB) narrowed the QRS duration to 90 ms, correcting the right conduction delay, while during left bundle branch block (LBBB), it produced broad complexes (140 ms) with fusion waves, failing to resolve the left-sided block. Right ventricular septal pacing may inadvertently correct RBBB but not LBBB, underscoring the importance of pacing site selection in patients with conduction abnormalities.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2026) conducted a case report in Superior vena cava occlusion and alternating bundle-branch block (n=1). Micra AV leadless pacemaker implantation (VDD pacing) was evaluated on QRS duration during VDD pacing. Micra AV VDD pacing narrowed the QRS duration to 90 ms during RBBB by correcting right conduction delay, but produced broad 140 ms complexes during LBBB.