Key result
The sensor-enhanced arthroscopic grasper demonstrated nearly perfect intraobserver agreement and substantial interobserver agreement for maximum load measurement of the supraspinatus tendon.
Why the study?
Objective data regarding rotator cuff tendon mobility are rare, leaving mobility to be evaluated subjectively by the surgeon.
A novel sensor-enhanced arthroscopic grasper provides reliable and objective measurements of supraspinatus tendon mobility in a porcine cadaver model.
Answers are grounded in this paper and cite their sources.
Supports objective tendon assessment in cadavers; leaves open human validation before clinical adoption.
Tendon mobility is highly relevant in rotator cuff surgery. Objective data about rotator cuff mobility is rare. Tendon mobility still needs to be evaluated subjectively by the surgeon. This study aims to establish a porcine animal model for mobility analysis of the supraspinatus. In this context, we introduce a sensor-enhanced, arthroscopic grasper (SEAG) suitable for objective intraoperative measurements of tendon mobility in clinical praxis. Tendon mobility of 15 fresh porcine cadaver shoulders with artificial rotator cuff tears was evaluated using the SEAG. Mobility characteristics (load-displacement curves, maximum load, stiffness) were studied and inter- and intraobserver agreement (intraclass correlation coefficient (ICC)) were tested. Factors with a potential adverse effect (plastic deformation and rigor mortis) were also evaluated. All shoulders showed characteristic reproducible load-displacement curves with a nonlinear part at the start, followed by a linear part. Mean maximum load was 28.6 N ± 12.5. Mean stiffness was 6.0 N/mm ± 2.6. We found substantial interobserver agreement (ICC 0.672) and nearly perfect intraobserver agreement (0.944) for maximum load measurement. Inter- (0.021) and intraobserver (0.774) agreement for stiffness was lower. Plastic deformation and rigor mortis were excluded. The animal model demonstrates reliable and in vivo-like measurements of tendon mobility. The SEAG is a reliable tool for tendon mobility assessment.
No takes yet. Share an insight, caveat, or question.
Porschke et al. (2020) studied Rotator cuff tear (artificial) (n=15). Sensor-enhanced, arthroscopic grasper (SEAG) was evaluated on Intraobserver agreement for maximum load measurement (ICC 0.944). The sensor-enhanced arthroscopic grasper demonstrated nearly perfect intraobserver agreement and substantial interobserver agreement for maximum load measurement of the supraspinatus tendon.