A low-friction surface is required for medical devices such as catheters or guidewires that are inserted into blood vessels. Failure of catheter procedures is often due to the inability to smooth insertion, followed by the subsequent vascular puncture at the tip. Although heparin contained saline can be lubricants among the two devices, there are also insert trouble when the surface contact pressure is high. Therefore, it is thought that the low friction between the catheter’s inner wall and guidewire is important, especially at bifurcation points and bends in blood vessels. The evaluation of the friction properties between guidewire and catheter (which material is PTFE) under different conditions (such as surface contact pressures and sliding speed) was conducted using an oscillating device, the effect of heparin on the tribological properties and lubrication states of guidewire against PTFE in saline solution was also evaluated. The film thickness ratio calculated based on material properties is 0.0001, indicating boundary lubrication conditions. The extremely low experimental friction coefficient suggests that under the lubrication conditions with heparin contained, the lowest friction coefficient of approximately 0.003 was achieved at 0.015 N and 0.2 mm/s, which the contact pressure is 2.9 MPa. It can be inferred that the addition of heparin formed a thicker lubrication film compared to saline in this condition, thereby reducing the friction coefficient.