Key result
Percutaneous delivery of 3D-printed coronary implants using a mother-and-child catheter was successfully achieved in 100% of swine models (11/11), with an 82% procedural survival rate.
Why the study?
Minimally invasive methods for creating focal coronary narrowing in large animals are challenging, and reliable percutaneous delivery of 3D printed coronary implants can be difficult without ancillary equipment.
Does the use of a mother-and-child coronary guide catheter enable reliable delivery of 3D printed coronary stenosis implants in swine models?
Does the use of a mother-and-child coronary guide catheter enable reliable delivery of 3D printed coronary stenosis implants in swine models?
Repurposing mother-and-child coronary guide catheters simplifies the percutaneous delivery of 3D printed coronary implants to create closed-chest swine models of focal coronary artery stenosis.
Answers are grounded in this paper and cite their sources.
Enhances large-animal coronary stenosis models; extends preclinical tools but leaves clinical translation open.
Minimally invasive methods for creating models of focal coronary narrowing in large animals are challenging. Rapid prototyping using three-dimensionally (3D) printed coronary implants can be employed to percutaneously create a focal coronary stenosis. However, reliable delivery of the implants can be difficult without the use of ancillary equipment. We describe the use of a mother-and-child coronary guide catheter for stabilization of the implant and for effective delivery of the 3D printed implant to any desired location along the length of the coronary vessel. The focal coronary narrowing was confirmed under coronary cineangiography and the functional significance of the coronary stenosis was assessed using gadolinium-enhanced first-pass cardiac perfusion MRI. We showed that reliable delivery of 3D printed coronary implants in swine models (n = 11) of ischemic heart disease can be achieved through repurposing mother-and-child coronary guide catheters. Our technique simplifies the percutaneous delivery of coronary implants to create closed-chest swine models of focal coronary artery stenosis and can be performed expeditiously, with a low procedural failure rate.
No takes yet. Share an insight, caveat, or question.
Hollowed et al. (2020) studied Ischemic heart disease (swine model) (n=11). Percutaneous delivery of 3D printed coronary stenosis implants using a mother-and-child catheter was evaluated on Successful delivery of the implant. Percutaneous delivery of 3D-printed coronary implants using a mother-and-child catheter was successfully achieved in 100% of swine models (11/11), with an 82% procedural survival rate.
Synapse has enriched 10 related papers. Consider them for comparative context: