Key result
The Low Frequency Attenuation (LFA) filter in St. Jude Medical ICDs can cause diaphragmatic myopotential oversensing, which can be resolved by deactivating the filter to avoid lead revision.
Why the study?
Does deactivating the Low Frequency Attenuation (LFA) filter resolve diaphragmatic myopotential oversensing in patients with St. Jude Medical ICDs?
Case Report
Does deactivating the Low Frequency Attenuation (LFA) filter resolve diaphragmatic myopotential oversensing in patients with St. Jude Medical ICDs?
Deactivating the Low Frequency Attenuation (LFA) filter in St. Jude Medical ICDs provides a simple programming solution to resolve diaphragmatic myopotential oversensing and avoid unnecessary lead revisions.
Answers are grounded in this paper and cite their sources.
May avoid lead revision by deactivating LFA filter in St. Jude ICDs; leaves open generalizability and durability.
Diaphragmatic myopotential oversensing (DMO) causes inhibition of pacing and inappropriate detection of ventricular fibrillation in implantable cardioverter defibrillators (ICDs). It occurs almost exclusively with integrated bipolar leads and is extremely rare with dedicated bipolar leads. If DMO cannot be corrected by reducing programmed sensitivity, ventricular lead revision is often required. The new Low Frequency Attenuation (LFA) filter in St. Jude Medical ICDs (St. Jude Medical, Sylmar, CA, USA) alters the sensing bandpass to reduce T-wave oversensing. This paper aims to present the LFA filter as a reversible cause of DMO. Unnecessary lead revision can be avoided by the simple programming solution of deactivating this LFA filter.
No takes yet. Share an insight, caveat, or question.
Ploux et al. (2016) conducted a case report in Diaphragmatic myopotential oversensing in implantable cardioverter defibrillators. Low Frequency Attenuation (LFA) filter was evaluated on Diaphragmatic myopotential oversensing (DMO). The Low Frequency Attenuation (LFA) filter in St. Jude Medical ICDs can cause diaphragmatic myopotential oversensing, which can be resolved by deactivating the filter to avoid lead revision.
Synapse has enriched 10 related papers. Consider them for comparative context: