Key result
Stored intracardiac electrograms confirmed pacemaker diagnostic counter triggers in 62% of episodes, revealing false-positive events in 385 cases.
Why the study?
Does the use of stored intracardiac electrograms accurately validate dual-chamber pacemaker diagnostic counter data in patients with standard pacemaker indications?
Observational (n=351)
Yes
Does the use of stored intracardiac electrograms accurately validate dual-chamber pacemaker diagnostic counter data in patients with standard pacemaker indications?
Stored intracardiac electrograms frequently reveal false-positive events in pacemaker diagnostic counters, highlighting the necessity of EGM validation before making therapeutic decisions.
May require EGM confirmation of pacemaker counters before clinical action; leaves open effects on outcomes in prospective studies.
BACKGROUND: Pacemaker diagnostic counters are used to guide device programming and patient management. However, these data are susceptible to inappropriate classification of events. The aim of this multicenter study was to evaluate pacemaker diagnostic data using stored intracardiac electrograms (EGMs). METHODS: The study included 351 patients (191 males, aged 71 +/- 10 years) with standard indications for dual-chamber pacemaker implantation. EGM triggers were atrial tachycardia (AT), ventricular tachycardia (VT), sudden bradycardia response (SBR), and pacemaker-mediated tachycardia (PMT). For this study, the devices could store up to 5 EGMs of 8s each (with marker annotation and onset recording). After 3 months, the EGMs were analyzed and classified as "confirmed" if the EGM validated the trigger and as "false positive" if the EGM showed an event different from the trigger. RESULTS: Of the 1,003 EGMs available, the triggers were AT in 640 EGMs, VT in 76, SBR in 105, and PMT in 178 EGMs. Four EGMs were triggered by magnet application. The trigger was confirmed in 614 EGMs (62%): 62% of AT episodes, 18% of VT episodes, 100% of SBR episodes, and 54% of PMT episodes. In 385 cases (45%), the EGMs revealed false-positive events due to far-field sensing (39%), noise and myopotential sensing (26%), sinus tachycardias (21%), double counting (9%), exit block (4%), and undersensing (1%). CONCLUSION: This large-scale study of stored EGMs revealed their value in validating diagnostic counter data. Therapeutic decisions should not be based on diagnostic counters alone; they should be validated by sophisticated tools like stored EGMs.
No takes yet. Share an insight, caveat, or question.
Nowak et al. (2005) conducted an observational in standard indications for dual-chamber pacemaker implantation (n=351). Stored intracardiac electrograms (EGMs) was evaluated on Confirmation of pacemaker diagnostic counter triggers. Stored intracardiac electrograms confirmed pacemaker diagnostic counter triggers in 62% of episodes, revealing false-positive events in 385 cases.