Key result
Telemetered atrial EGM corrects surface ECG misdiagnosis by revealing far-field sensing of ventricular depolarization.
Why the study?
Does telemetered atrial electrogram improve the diagnosis of pacemaker-mediated arrhythmias compared to surface electrocardiography in patients with dual-chamber pacemakers?
Case Report (n=1)
Does telemetered atrial electrogram improve the diagnosis of pacemaker-mediated arrhythmias compared to surface electrocardiography in patients with dual-chamber pacemakers?
Telemetered atrial electrograms are valuable for accurately diagnosing pacemaker-mediated arrhythmias that may be misdiagnosed on surface ECG.
Answers are grounded in this paper and cite their sources.
May clarify ECG-misdiagnosed pacemaker arrhythmias; leaves open prospective validation in dual-chamber devices.
A 73-year-old male required a permanent pacemaker for complete heart block. After placement of the dual-chamber pacemaker, the patient developed random, repetitive ventricularly paced coupling at the upper rate limit. The diagnosis made from the surface electrocardiography was one of ventricular tracking of occult (T wave obscured) atrial ectopic activity. Examination of the telemetered atrial electrogram (EGM) revealed a diagnosis of intermittent far-field atrial sensing of the terminal ventricular depolarization. The value of obtaining an atrial electrogram during dual-chamber pacing is discussed.
No takes yet. Share an insight, caveat, or question.
Shandling et al. (1988) conducted a case report in Complete heart block and pacemaker-mediated arrhythmia (n=1). Telemetered atrial electrogram (EGM) vs. Surface electrocardiography was evaluated on Diagnosis of pacemaker-mediated arrhythmia. Telemetered atrial electrogram revealed intermittent far-field atrial sensing of terminal ventricular depolarization, correcting the misdiagnosis of ventricular tracking made by surface ECG.
Synapse has enriched 10 related papers. Consider them for comparative context: