Key result
A variable-rate selective excitation (VERSE)-guided numerical RF pulse design framework was developed to simultaneously control peak RF power and compensate for off-resonance in parallel excitation.
The study proposes a VERSE-guided numerical RF pulse design to control peak RF power and compensate for off-resonance in parallel excitation MRI.
May reduce peak RF power in parallel excitation MRI; leaves open prospective clinical validation.
In parallel excitation, the computational speed of numerical radiofrequency (RF) pulse design methods is critical when subject dependencies and system nonidealities need to be incorporated on-the-fly. One important concern with optimization-based methods is high peak RF power exceeding hardware or safety limits. Hence, online controllability of the peak RF power is essential. Variable-rate selective excitation pulse reshaping is ideally suited to this problem due to its simplicity and low computational cost. In this work, we first improve the fidelity of variable-rate selective excitation implementation for discrete-time waveforms through waveform oversampling such that variable-rate selective excitation can be robustly applied to numerically designed RF pulses. Then, a variable-rate selective excitation-guided numerical RF pulse design is suggested as an online RF pulse design framework, aiming to simultaneously control peak RF power and compensate for off-resonance.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2011) studied this question. VERSE-guided numerical RF pulse design was evaluated on Peak RF power control and off-resonance compensation. A variable-rate selective excitation (VERSE)-guided numerical RF pulse design framework was developed to simultaneously control peak RF power and compensate for off-resonance in parallel excitation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: