High-resolution ultrasonic imaging needs precise control over acoustic field uniformity and beam profile features to reduce imaging errors and improve spatial resolution. Limitations like sidelobe interference and uneven pressure continue to hinder system performance in advanced imaging applications. This study aims to analyze acoustic field distribution and beam behavior under various transducer setups and excitation conditions. We used finite-element acoustic simulations to assess spatial pressure fields, beam width, sidelobe suppression, focal stability, and lateral resolution throughout the imaging area. We compared conventional and optimized transducer designs that operated at the same frequencies and boundary conditions. The optimized design showed a 32% to 38% increase in field uniformity while reducing sidelobe amplitude by about 41% compared to the baseline design. Lateral resolution at the focal zone improved from 1.12 mm to 0.76 mm, with a 29% increase in peak pressure stability along the beam axis. These results show that focused transducer optimization greatly enhances acoustic beam performance, leading to clearer images and less distortion. The findings offer useful design insights for future ultrasonic imaging and nondestructive evaluation systems. These improvements directly enhance signal retrieval and image reconstruction quality by improving contrast, edge definition, and noise robustness in high-frequency ultrasound imaging.