Key result
Transthoracic impedance measured via tripolar ICD leads correlated well with pneumotachometer measurements for minute ventilation (r=0.86 ± 0.07), suggesting feasibility for MV sensor systems.
Why the study?
Can ICD leads be used accurately as part of a minute ventilation detection system compared to a pneumotachometer?
Observational (n=10)
Can ICD leads be used accurately as part of a minute ventilation detection system compared to a pneumotachometer?
Tripolar and tetrapolar ICD lead vectors provide accurate minute ventilation and respiration rate measurements, suggesting their feasibility for use in MV sensor systems for rate-responsive pacing.
May enable MV-based rate response via existing ICD leads; hypothesis-generating pending prospective validation.
BACKGROUND: Minute ventilation (MV) has been proven to be very useful in rate responsive pacing. The aim of this study was to evaluate the feasibility of using implantable cardioverter-defibrillator (ICD) leads as part of the MV detection system. METHODS: At implant in 10 patients, the transthoracic impedance was measured from tripolar ICD, tetrapolar ICD, and atrial lead vectors during normal, deep, and shallow voluntary respiration. MV and respiration rate (RespR) were simultaneously measured through a facemask with a pneumotachometer (Korr), and the correlations with impedance-based measurements were calculated. Air sensitivity was the change in impedance per change in respiratory tidal volume, ohms (Omega)/liter (L), and the signal-to-noise ratio (SNR) was the ratio of the respiratory and cardiac contraction components. RESULTS: The air sensitivity and SNR in tripolar ICD vector were 2.70 +/- 2.73 ohm/L and 2.19 +/- 1.31, respectively, and were not different from tetrapolar. The difference in RespR between tripolar ICD and Korr was 0.2 +/- 1.91 breaths/minute. The regressed correlation coefficient between impedance MV and Korr MV was 0.86 +/- 0.07 in tripolar ICD. CONCLUSIONS: The air sensitivity and SNR in tripolar and tetrapolar ICD lead vectors did not differ significantly and were in the range of the values in pacemaker leads currently used as MV sensors. The good correlations between impedance-based and Korr-based RespR and MV measurements imply that ICD leads may be used in MV sensor systems.
No takes yet. Share an insight, caveat, or question.
Gottfridsson et al. (2009) conducted an observational in Patients undergoing ICD implantation (n=10). Transthoracic impedance measurement using ICD leads vs. Pneumotachometer (Korr) was evaluated on Correlation between impedance-based and pneumotachometer-based minute ventilation. Transthoracic impedance measured via tripolar ICD leads correlated well with pneumotachometer measurements for minute ventilation (r=0.86 ± 0.07), suggesting feasibility for MV sensor systems.