Key result
Pacemaker flipping in a single chamber activity sensing rate responsive pacemaker caused pocket pacing and positive feedback, resulting in pacemaker-mediated tachycardia at rest up to 106 beats/min.
Case Report (n=1)
Pacemaker flipping in single chamber activity sensing rate responsive pacemakers can cause pacemaker mediated tachycardia due to pocket pacing activating the sensor.
May cause unexplained resting tachycardia in activity-sensing pacemakers; leaves open incidence and prevention in larger cohorts.
Although pacemaker mediated tachycardias are classically associated with dual chamber pacemakers, single chamber rate responsive pacemakers are also susceptible to such tachycardias under special circumstances. A unipolar activity sensing rate responsive pacemaker (Activitrax 8403) was implanted in an 83-year-old man with complete atrioventricular block. The pacemaker was programmed at an output of 5 V, activity threshold medium, rate response 5, and lower and upper rates of 70 and 125 beats/min, respectively. He presented with palpitations at rest and muscle twitching of the pacemaker pocket 4 months after implantation. Examination confirmed that the pacemaker had flipped over, resulting in pocket pacing which in turn activated the activity sensor, resulting in a rate response. The increase in pacing rate lead to a higher frequency of pocket pacing, thus leading to positive feedback increase in rate. With the patient at rest, pacemaker mediated rates were 106, 91, and 74 beats/min at low, medium and high thresholds, respectively. Decreasing the output to 2.5 V eliminated pocket pacing and the tachycardia. As a result of the reversal of the pacemaker, a similar rate response during exercise could only be achieved at a more sensitive rate responsive setting. Thus, pacemaker mediated tachycardia can complicate pacemaker "flipping" in single chamber activity sensing rate responsive pacemakers. Methods for the avoidance and treatment of pacemaker flipping are discussed. A review of other sensor mediated tachycardias is also presented.
No takes yet. Share an insight, caveat, or question.
Lau et al. (1990) conducted a case report in Complete atrioventricular block (n=1). Unipolar activity sensing rate responsive pacemaker (Activitrax 8403) was evaluated on Pacemaker mediated tachycardia. Pacemaker flipping in a single chamber activity sensing rate responsive pacemaker caused pocket pacing and positive feedback, resulting in pacemaker-mediated tachycardia at rest up to 106 beats/min.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: