Key result
Dynamic 40-slice ECG-gated cine CTA demonstrated a 2.2- to 3.4-mm increase in aortic wall circumference and a 2.4- to 3.6-mm increase in luminal circumference per cardiac cycle in AAA patients.
Why the study?
Does 40-slice ECG-gated cine CTA accurately characterize aortic motion and circumference changes during the cardiac cycle in preoperative AAA patients?
Observational (n=10)
Does 40-slice ECG-gated cine CTA accurately characterize aortic motion and circumference changes during the cardiac cycle in preoperative AAA patients?
Dynamic cine CTA demonstrates significant changes in aortic circumference during the cardiac cycle in AAA patients, which may have important implications for endograft sizing and design.
Answers are grounded in this paper and cite their sources.
Dynamic aortic expansion in AAA may affect endograft sizing; hypothesis-generating and should not yet change practice.
PURPOSE: To utilize 40-slice electrocardiographically (ECG)-gated cine computed tomographic angiography (CTA) to characterize normal aortic motion during the cardiac cycle at relevant anatomical landmarks in preoperative abdominal aortic aneurysm (AAA) patients. METHODS: In 10 consecutive preoperative AAA patients (10 men; mean age 78.8 years, range 69-86), an ECG-gated CTA dataset was acquired on a 40-slice CT scanner using a standard radiation dose. CTA quality was graded and scan time was measured. Pulsatility measurements at multiple relevant anatomical levels were performed in the axial plane. Changes in aortic circumference were determined for both the aortic wall and the luminal diameter. RESULTS: All 10 CT scans were of good quality. All patients could be scanned in 14 to 33 seconds (mean 21). At each anatomical level measured, there was a 2.2- to 3.4-mm increase in the aortic wall circumference per cardiac cycle. A similar increase was observed in luminal circumference, with a 2.4- to 3.6-mm increase per cycle. CONCLUSION: This study introduces the concept of dynamic cine CTA imaging of aortic motion, providing insight into the pathophysiology of abdominal aortic and iliac pulsations. Patients with AAAs selected for EVAR demonstrate changes in aortic circumference with each cardiac cycle that may have consequences for endograft sizing and future design. The potential for graft migration, intermittent type I endoleak, and poor patient outcome following EVAR can be anticipated. Complex aortic dynamics deserve increased scrutiny in an effort to prevent potential complications.
No takes yet. Share an insight, caveat, or question.
Teutelink et al. (2006) conducted an observational in abdominal aortic aneurysm (n=10). 40-slice ECG-gated cine computed tomographic angiography (CTA) was evaluated on Changes in aortic circumference for both the aortic wall and the luminal diameter. Dynamic 40-slice ECG-gated cine CTA demonstrated a 2.2- to 3.4-mm increase in aortic wall circumference and a 2.4- to 3.6-mm increase in luminal circumference per cardiac cycle in AAA patients.
Synapse has enriched 10 related papers. Consider them for comparative context: