Key result
In hemodialysis patients with PTFE grafts, an access blood flow of less than 500 ml/min tripled the risk of subsequent thrombosis compared to a flow of 1100-1400 ml/min.
Why the study?
Does repeated measurement of access blood flow predict access thrombosis in patients on hemodialysis?
Cohort (n=131)
Does repeated measurement of access blood flow predict access thrombosis in patients on hemodialysis?
Answers are grounded in this paper and cite their sources.
Repeated measurement of access blood flow can predict future access failure in hemodialysis patients with PTFE grafts, but this relationship was not observed in AV fistulas.
The aim of this study was to evaluate whether repeated measurement of access blood flow (Qac) using the ultrasound dilution technique could predict access failure in patients on hemodialysis. One hundred thirty-one patients were evaluated at intervals of 8 weeks for a period of 6 months. The incidence of thrombosis was determined within each study period. During the 6 month follow-up, 36 thrombotic events occurred in 27 of 68 polytetrafluoroethylene (PTFE) grafts, and only six thrombotic events in 5 of 63 arteriovenous (AV) fistulas. The relative risk for access thrombosis for patients with PTFE grafts was 5.6 times greater than for patients with AV fistulas. Qac was significantly lower in thrombotic compared with patent PTFE grafts (958 +/- 506 ml/min vs 1141 +/- 482 ml/min, p < 0.05). A significant relationship was found between the incidence of subsequent PTFE graft thrombotic events and Qac (p < 0.001). Compared with accesses with high blood flow (1100-1400 ml/min), the risk for subsequent thrombosis tripled in grafts with a Qac of less than 500 ml/min. This relationship was not seen with AV fistulas. In patent PTFE grafts, Qac remained unchanged within each 2 month interval, whereas it decreased in thrombotic PTFE grafts. Thus, repeated measurements of Qac have the potential to predict future access failure in PTFE grafts; however, an increased measuring frequency might improve the predictive value of graft failure with high Qac.
No takes yet. Share an insight, caveat, or question.
Wang et al. (1998) conducted a cohort in Hemodialysis (n=131). Access blood flow (Qac) < 500 ml/min vs. Access blood flow 1100-1400 ml/min was evaluated on Access thrombosis (RR 3.0, p=<0.001). In hemodialysis patients with PTFE grafts, an access blood flow of less than 500 ml/min tripled the risk of subsequent thrombosis compared to a flow of 1100-1400 ml/min.