Answers are grounded in this paper and cite their sources.
Serum phosphorus can be measured by continuous-flow uv spectrophotometry without reduction of the phosphomolybdate complex. The dilute sample is dialyzed into dilute (1 ml/100 ml) sulfuric acid, then mixed with an ammonium molybdate— sulfuric acid—"Tween 80" solution. The absorbance of the sample peaks is measured at 340 nm with a linear-absorbance spectrophotometer. Peak heights are directly proportional to concentration, because logarithmic conversion is performed within the spectrophotometer. The method obeys Beer’s law up to 10 mg of P per deciliter, and results correlate closely with those for standard methods based on reduced phosphomolybdate blue. The miniature manifold uses a 12-in. dialyzer. The sampling rate can be 60 to 120 samples per hour when a Gilford one-piece debubbler flow cell is used.
No takes yet. Share an insight, caveat, or question.
Amador et al. (1972) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: