Key result
In a mock extracorporeal system, higher postpump resistance reduced pump flow range and increased postpump pressure and surplus hemodynamic energy output with pulsatile mode.
Why the study?
Does pulsatile mode compared to nonpulsatile mode alter pressure-flow waveforms and hemodynamic energy output under varying postpump resistances in a neonatal centrifugal pump mock circuit?
Does pulsatile mode compared to nonpulsatile mode alter pressure-flow waveforms and hemodynamic energy output under varying postpump resistances in a neonatal centrifugal pump mock circuit?
In a neonatal mock circuit, higher postpump resistance reduces pump flow range but increases surplus hemodynamic energy output during pulsatile centrifugal pump operation.
Answers are grounded in this paper and cite their sources.
Caution against clinical adoption in neonates; leaves open optimal resistance settings for pulsatile centrifugal pumps pending in vivo data.
This study tested the impact of different postpump resistances on pulsatile pressure-flow waveforms and hemodynamic energy output in a mock extracorporeal system. The circuit was primed with a 40% glycerin-water mixture, and a PediVAS centrifugal pump was used. The pre- and postpump pressures and flow rates were monitored via a data acquisition system. The postpump resistance was adjusted using a Hoffman clamp at the outlet of the pump. Five different postpump resistances and rotational speeds were tested with nonpulsatile (NP: 5000 RPM) and pulsatile (P: 4000 RPM) modes. No backflow was found when using pulsatile flow. With isoresistance, increased arterial resistances decreased pump flow rates (NP: from 1,912 ml/min to 373 ml/min; P: from 1,485 ml/min to 288 ml/min), increased postpump pressures (NP: from 333 mm Hg to 402 mm Hg; P: from 223 mm Hg to 274 mm Hg), and increased hemodynamic energy output with pulsatile mode. Pump flow rate correlated linearly with rotational speed (RPMs) of the pump, whereas postpump pressures and hemodynamic energy outputs showed curvilinear relationships with RPMs. The maximal pump flow rate also increased from 618 ml/min to 4,293 ml/min with pulsatile mode and from 581 ml/min to 5,665 ml/min with nonpulsatile mode. Results showed that higher postpump resistance reduced the pump flow range, and increased postpump pressure and surplus hemodynamic energy output with pulsatile mode. Higher rotational speeds also generated higher pump flow rates, postpump pressures, and increased pulsatility.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2009) studied this question. Pulsatile mode with varying postpump resistances vs. Nonpulsatile mode (5000 RPM) was evaluated on Pump flow rates, postpump pressures, and hemodynamic energy output. In a mock extracorporeal system, higher postpump resistance reduced pump flow range and increased postpump pressure and surplus hemodynamic energy output with pulsatile mode.