A study of patient exposure in paediatric diagnostic radiology was conducted in three dedicated paediatric departments. The entrance surface dose was assessed by three methods: direct measurement by thermoluminescent dosemeters, calculation from the dose-area product and from the tube output. The results obtained by the these methods were compared, proving that all of them are applicable for the assessment of paediatric doses. Subsequently, the accumulated data were compared with the mean values from other similar studies and to the reference dose levels set by the Commission of European Communities, which clearly demonstrated the need for further investigation of the paediatric dose levels in Bulgaria and for optimisation of the radiological practice.