Key result
In an ex vivo mock loop, Trifecta showed lower pressure drop and greater effective orifice area at stroke volumes of 30-60 ml, but Mitroflow slightly outperformed Trifecta at higher stroke volumes.
Why the study?
Does the Trifecta bioprosthesis improve hydrodynamic performance compared to the Mitroflow bioprosthesis in an ex vivo model of small aortic roots?
Does the Trifecta bioprosthesis improve hydrodynamic performance compared to the Mitroflow bioprosthesis in an ex vivo model of small aortic roots?
The hydrodynamic performance of bioprostheses in small aortic roots depends on complex geometric behaviors at different stroke volumes, not just internal diameter.
Answers are grounded in this paper and cite their sources.
May guide valve selection by stroke volume in small roots; leaves open clinical translation pending in vivo data.
Hydrodynamic performance of stented bioprostheses is far below that of the native valve. One of the reasons is that the internal diameter of the prosthesis is usually smaller than that of the native valve. However, other valve characteristics are also important in generating the pressure drop. We aimed to assess, in an ex vivo pulsatile mock loop, the hydrodynamic behavior of two bioprostheses, Trifecta and Mitroflow, to ascertain which geometric terms are limiting factors in hydrodynamic performance. At stroke volumes between 30 and 60 ml, Trifecta showed lower pressure drop, energy dissipation and valve resistance, and greater effective orifice area. This trend was overturned at higher stroke volumes, with Mitroflow slightly outperforming Trifecta. The geometric determinants were consistent with these results. Trifecta achieved its maximum opening area already at the lowest stroke volumes, featuring a divergent shape at the systolic peak. Mitroflow showed a complex opening pattern, featuring a convergent shape at the systolic peak for lower stroke volumes, while reaching its maximum opening area at higher stroke volumes, with a divergent shape. The two bioprostheses, although similar in design, displayed different biomechanical behaviors. The internal diameter of each bioprosthesis did not show to be strictly correlated with its hydrodynamic characteristics.
No takes yet. Share an insight, caveat, or question.
Tasca et al. (2017) studied Small Aortic Roots. Trifecta bioprosthesis vs. Mitroflow bioprosthesis was evaluated on Hydrodynamic behavior (pressure drop, energy dissipation, valve resistance, effective orifice area). In an ex vivo mock loop, Trifecta showed lower pressure drop and greater effective orifice area at stroke volumes of 30-60 ml, but Mitroflow slightly outperformed Trifecta at higher stroke volumes.