Key result
A collapsible inflow cannula reservoir minimized suction events and maintained positive chamber volumes compared to a rigid cannula in an in vitro mock circulation loop.
Why the study?
Does a collapsible inflow cannula reservoir prevent suction events compared to a rigid cannula in an in vitro mock circulation loop of extracorporeal rotary LVAD support?
Does a collapsible inflow cannula reservoir prevent suction events compared to a rigid cannula in an in vitro mock circulation loop of extracorporeal rotary LVAD support?
A collapsible inflow cannula reservoir effectively minimizes suction events in an in vitro model of extracorporeal rotary LVAD support.
Answers are grounded in this paper and cite their sources.
May inform LVAD cannula design to reduce suction; leaves open translation to clinical use.
Limited preload sensitivity of rotary left ventricular assist devices (LVADs) renders patients susceptible to harmful atrial or ventricular suction events. Active control systems may be used to rectify this problem; however, they usually depend on unreliable sensors or potentially inaccurate inferred data from, for example, motor current. This study aimed to characterize the performance of a collapsible inflow cannula reservoir as a passive control system to eliminate suction events in extracorporeal, rotary LVAD support. The reservoir was evaluated in a mock circulation loop against a rigid cannula under conditions of reduced preload and increased LVAD speed in both atrial and ventricular cannulation scenarios. Both cases demonstrated the ease with which chamber suction events can occur with a rigid cannula and confirm that the addition of the reservoir maintained positive chamber volumes with reduced preload and high LVAD speeds. Reservoir performance was dependent on height with respect to the cannulated chamber, with lower placement required in atrial cannulation due to reduced filling pressures. This study concluded that a collapsible inflow cannula is capable of minimizing suction events in extracorporeal, rotary LVAD support.
No takes yet. Share an insight, caveat, or question.
Gregory et al. (2011) studied Extracorporeal rotary ventricular assist device support. Collapsible inflow cannula reservoir vs. Rigid cannula was evaluated on Suction events and chamber volumes. A collapsible inflow cannula reservoir minimized suction events and maintained positive chamber volumes compared to a rigid cannula in an in vitro mock circulation loop.