Stent grafts have been used for a variety of arteriovenous access associated issues. This article presents three cases of stent graft infection and a case of protruded metal piece of the stent graft through the skin. All four required surgical treatment and three cases required a tunneled dialysis catheter to provide long-term dialysis therapy. This report highlights that stent graft problems can occur that may result in loss of the access. Additionally, strut protrusion can pose a medical hazard to those performing preparation and cannulation of the arteriovenous access.
No takes yet. Share an insight, caveat, or question.
Asif et al. (2010) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: