This is the authors' abstract. We don't add key points for this paper.
BACKGROUND: Intramedullary screw fixation of Jones fractures using partially threaded screws is a common method of fixation for these injuries, but refracture continues to be a problem. Various other fixation strategies, such as headless compression screws, plantar plating, and tension-band wiring. have been developed to mitigate these issues. Biomechanical studies with regard to these other fixation strategies are limited. Herein, we investigate the compression strength and angular stiffness of Jones fractures fixed with Herbert-style headless compression screws. METHODS: Jones fractures were created in 10 fresh-frozen pairs of cadaveric fifth metatarsals. A bone from each pair was instrumented with either a conventional, partially threaded screw 5.0 or 6.5 mm in diameter, or a headless compression screw 5.0 or 7.0 mm in diameter. Sizes were determined via sequential tapping until a snug fit was obtained. Each metatarsal was stressed via cantilever bending over 1000 cycles. We monitored compression and displacement throughout. RESULTS: Headless compression screws achieved a significantly higher amount of stiffness than conventional, partially threaded screws (P = 0.005). There was no statistically significant difference with respect to compression. CONCLUSION: Therapeutic, Level V: Biomechanical.