Key result
Automatic switching between AAI and DDD pacing modes prevented repetitive non-reentrant ventriculoatrial synchrony in a patient with an implantable cardioverter-defibrillator.
Why the study?
Does automatic switching between AAI and DDD algorithms prevent repetitive non-reentrant ventriculoatrial synchrony in patients with ICDs?
Case Report (n=1)
Does automatic switching between AAI and DDD algorithms prevent repetitive non-reentrant ventriculoatrial synchrony in patients with ICDs?
Automatic switching between AAI and DDD algorithms in ICDs may effectively prevent repetitive non-reentrant ventriculoatrial synchrony by avoiding unnecessary ventricular pacing.
May prevent repetitive non-reentrant ventriculoatrial synchrony in ICD patients via AAI-DDD switching; leaves open prospective validation.
A 67‐year‐old man with non‐obstructive hypertrophic cardiomyopathy had received an implantable cardioverter‐defibrillator (ICD) for an unstable, sustained ventricular tachycardia (VT) induced by programmed stimulation during an electrophysiological study 5 years earlier. An intracardiac electrogram recorded by the ICD revealed repetitive, non‐reentrant ventriculoatrial synchrony (RNRVAS) associated with hypotension. Electrophysiologic and hemodynamic studies indicated that RNRVAS was induced and reproducibly termed by a single ventricular extrastimulus from the right ventricular apex. Following attainment of the elective replacement indicator, we replaced the ICD with another having managed ventricular pacing, which automatically switched AAI and DDD, thereby avoiding unnecessary ventricular pacing. Thus far, the patient has not experienced further RNRVAS. Thus, we believe that automatic switching between AAI and DDD can prevent RNRVAS.
No takes yet. Share an insight, caveat, or question.
Kitamura et al. (2013) conducted a case report in Non-obstructive hypertrophic cardiomyopathy with repetitive non-reentrant ventriculoatrial synchrony (n=1). Automatic switching between AAI and DDD (managed ventricular pacing) was evaluated on Prevention of repetitive non-reentrant ventriculoatrial synchrony (RNRVAS). Automatic switching between AAI and DDD pacing modes prevented repetitive non-reentrant ventriculoatrial synchrony in a patient with an implantable cardioverter-defibrillator.