Key result
Valve-in-valve TAVI with Portico or Navitor valves for failed surgical bioprostheses achieved 89.4% technical success and favorable 30-day safety outcomes with no deaths.
Why the study?
Valve-in-valve TAVI is an established alternative to redo surgical replacement for degenerate bioprostheses, but data on the performance of self-expanding Portico and Navitor valves remain limited.
Observational (n=47)
No
ViV TAVI using Portico or Navitor valves demonstrates excellent short-term safety and technical success for treating failed surgical bioprostheses, including in patients with small surgical valves.
BACKGROUND: Degeneration of surgical bioprosthetic aortic valves is increasingly common. Redo surgical aortic valve replacement carries substantial morbidity and mortality, particularly in elderly or high-risk patients. Valve-in-valve (ViV) transcatheter aortic valve implantation (TAVI) has become an established alternative, though data on the performance of self-expanding Portico and Navitor valves remain limited. METHODS: We retrospectively analyzed 47 consecutive patients who underwent ViV TAVI for severe bioprosthetic aortic valve dysfunction between January 2020 and June 2024 at Barts Heart Center. In-hospital and 30-day outcomes were evaluated using Valve Academic Research Consortium 3 criteria. RESULTS: The median age was 79 years, and 80.9% had small surgical valves (≤ 23 mm). ViV TAVI with Portico or Navitor valves was successful in 46 cases (97.9%), with no in-hospital mortality or need for emergency cardiac surgery. Technical success was achieved in 89.4% of cases; failures were related to two major vascular complications, two valve-related interventions, and one major cardiac structural complication. At 30 days, device success and safety outcomes were favorable, with no deaths or additional major complications. Two patients (4.3%) experienced ischemic stroke, and one (2.8%) required a new permanent pacemaker. Most patients (97.8%) were in NYHA class I-II. Echocardiography confirmed intended valve performance in 87% of cases, with no moderate or severe aortic regurgitation. CONCLUSIONS: ViV TAVI with Portico and Navitor valves is a safe and effective option for failed surgical bioprostheses, including small valves, offering excellent short-term outcomes. Further studies are needed to assess long-term durability.
No takes yet. Share an insight, caveat, or question.
Tufaro et al. (2025) conducted an observational in severe bioprosthetic aortic valve dysfunction (n=47). Valve-in-valve TAVI with Portico or Navitor valves was evaluated on Technical success (Valve Academic Research Consortium 3 criteria). Valve-in-valve TAVI with Portico or Navitor valves for failed surgical bioprostheses achieved 89.4% technical success and favorable 30-day safety outcomes with no deaths.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: