Key result
Valve-in-valve TAVR using the Portico/Navitor intra-annular self-expanding valve achieved a 30-day device success rate of 73.5%, with a 2-year survival rate of 85.6%.
Why the study?
Data on intra-annular self-expanding valves for valve-in-valve TAVR remain limited compared to balloon-expandable and supra-annular self-expanding valves.
Observational (n=200)
Yes
Valve-in-valve TAVR using the intra-annular self-expanding Portico/Navitor systems demonstrated a 73.5% 30-day device success rate and favorable 2-year survival in a high-risk cohort with predominantly small surgical bioprostheses.
May guide valve selection for high-risk valve-in-valve TAVR in small bioprostheses; extends observational data to intra-annular self-expanding platforms.
BACKGROUND: Most evidence for transcatheter aortic valve replacement (TAVR) for valve-in-valve procedures comes from balloon-expandable and supra-annular self-expanding valves. However, data on intra-annular self-expanding valves remain limited. The objectives of this registry were to evaluate the short- and midterm outcomes of ViV-TAVR using the intra-annular self-expanding valve Navitor platform in a real-world setting. METHODS: The NAViV (Navitor Valve-in-Valve) registry is a retrospective, multicenter study that included 200 patients who underwent valve-in-valve TAVR with the Portico/Navitor systems. The primary end point was 30-day device success. Secondary outcomes included intraprocedural technical success, early safety, and all-cause death and major adverse events at 2 years. RESULTS: The cohort had a high-risk profile and a high prevalence of small surgical bioprostheses (internal diameter ≤21 mm in 86% of the patients). Technical success was achieved in 96%, with an in-hospital mortality rate of 2% and early safety at 30 days of 84%. Mean residual gradients after TAVR were low (14 [interquartile range, 10-20] mm Hg), and residual significant aortic regurgitation was rare (1%). Device success was 73.5%, with high residual gradients (≥20 mm Hg) being the main reason for device failure. Two-year survival and major adverse events were 85.6% and 69.8%, respectively; with low residual gradients and baseline significant AR being independently associated with higher survival. CONCLUSIONS: Valve-in-valve TAVR using the Portico/Navitor intra-annular self-expanding valve demonstrates a favorable procedural safety profile, sustained valve performance, and encouraging midterm survival in intermediate- and high-risk populations. Larger and comparative studies are warranted to confirm these findings.
No takes yet. Share an insight, caveat, or question.
Solsol et al. (2026) conducted an observational in Failing surgical aortic bioprostheses (n=200). Valve-in-valve TAVR with Portico/Navitor systems was evaluated on 30-day device success. Valve-in-valve TAVR using the Portico/Navitor intra-annular self-expanding valve achieved a 30-day device success rate of 73.5%, with a 2-year survival rate of 85.6%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: