Key result
The finite state machine approach successfully simulated changes in preload, afterload, and contractility with an error of less than 1 mmHg and 1 mL between desired and simulated set points.
Why the study?
Historical left ventricle models based on biomechanical analysis and continuous states are not conducive to deterministic processing, prompting the need for finite-state machine governance.
The proposed finite state machine approach accurately simulates left ventricular pressure-volume relationships, enabling personalized in silico modeling and deterministic control for mock circulatory systems.
Facilitates precise in silico LV modeling for research; leaves open human validation before clinical mock circulatory use.
BACKGROUND: Simulation of a left ventricle has become a critical facet of evaluating therapies and operations that interact with cardiac performance. The ability to simulate a wide range of possible conditions, changes in cardiac performance, and production of nuisances at transition points enables evaluation of precision medicine concepts that are designed to function through this spectrum. Ventricle models have historically been based on biomechanical analysis, with model architectures constituted of continuous states and not conducive to deterministic processing. Producing a finite-state machine governance of a left ventricle model would enable a broad range of applications: physiological controller development, experimental left ventricle control, and high throughput simulations of left ventricle function. METHODS: and Stateflow tools. RESULTS: The approach developed is capable of simulating changes in preload, afterload, and contractility in time based on a patient's preload analysis. Six pressure-volume loop simulations are presented to include a base-line, preload change only, afterload change only, contractility change only, a clinical control, and heart failure with normal ejection fraction. All simulations produced an error of less than 1 mmHg and 1 mL of the absolute difference between the desired and simulated pressure and volume set points. The acceptable performance of the fixed-timestep architecture in the finite state machine allows for deployment to deterministic systems, such as experimental systems for validation. CONCLUSIONS: The proposed approach allows for personalized data, revealed through an individualized clinical pressure-volume analysis, to be simulated in silico. The computational model architecture enables this control structure to be executed on deterministic systems that govern experimental left ventricles. This provides a mock circulatory system with the ability to investigate the pathophysiology for a specific individual by replicating the exact pressure-volume relationship defined by their left ventricular functionality; as well as perform predictive analysis regarding changes in preload, afterload, and contractility in time.
No takes yet. Share an insight, caveat, or question.
King et al. (2019) studied Left Ventricular Modeling. Finite state machine implementation for left ventricle modeling vs. Desired theoretical set points was evaluated on Absolute difference between desired and simulated pressure and volume set points. The finite state machine approach successfully simulated changes in preload, afterload, and contractility with an error of less than 1 mmHg and 1 mL between desired and simulated set points.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: