Key result
A finite element driven design domain identification demonstrated that a beating left ventricular simulator configuration with 4 mm wall thickness and 8 actuators at 90 degrees orientation performed best.
Why the study?
Existing beating left ventricular simulators mimicking heart wall motion had flow rates limited to 2 liters per minute.
A finite element model identified optimal design parameters (4 mm wall thickness, 8 actuators, 90-degree orientation) to improve the deformation of a beating left ventricular simulator for LVAD testing.
Almost ten percent of the American population have heart diseases. Since the number of available heart donors is not promising, left ventricular assist devices are implemented as bridge therapies. Development of the assist devices benefits from both in-vivo animal and in-vitro mock circulation studies. Representation of the heart is a crucial part of the mock circulation setups. Recently, a beating left ventricular simulator with latex rubber and helically oriented McKibben actuators has been proposed. The simulator was able to mimic heart wall motion, however, flow rate was reported to be limited to 2 liters per minute. This study offers a finite element driven design domain identification to identify the combination of wall thickness, number of actuators, and the orientation angle that results in better deformation. A nonlinear finite element model of the simulator was developed and validated. Design domain was constructed with 150 finite element models, each with varying wall thickness and number of actuators with varying orientation angles. Results showed that the combination of 4 mm wall thickness and 8 actuators with 90 degrees orientation performed best in the design domain.
No takes yet. Share an insight, caveat, or question.
Gülbulak et al. (2019) studied Heart diseases (Left ventricular simulator design). Beating left ventricular simulator with varying wall thickness and actuator configurations was evaluated on Directional deformation and rotation of the apex. A finite element driven design domain identification demonstrated that a beating left ventricular simulator configuration with 4 mm wall thickness and 8 actuators at 90 degrees orientation performed best.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: