Key result
Transcatheter tricuspid valve replacement using the LuX-Valve with 3D printing guidance successfully reduced postprocedural tricuspid regurgitation in a high-risk patient.
Why the study?
Transcatheter treatments for tricuspid regurgitation have emerged for high-risk patients, with radial force-independent stent valve devices offering distinct advantages at the native tricuspid annular site.
Does transcatheter tricuspid valve replacement using the LuX-Valve with 3D printing guidance reduce tricuspid regurgitation in a high-risk patient with prior valve replacements?
Case Report (n=1)
Does transcatheter tricuspid valve replacement using the LuX-Valve with 3D printing guidance reduce tricuspid regurgitation in a high-risk patient with prior valve replacements?
Transcatheter tricuspid valve replacement using the LuX-Valve with 3D printing guidance is feasible and safe in a high-risk patient with prior mechanical mitral and transcatheter aortic valves.
BACKGROUND: Transcatheter treatments of tricuspid regurgitation (TR) have been emerging as alternatives for high-risk patients. In contrast to the immobilization of the common transcatheter tricuspid device, using a radial force-independent stent valve device at the native tricuspid annular site has several distinct advantages. CASE SUMMARY: A 76-year-old patient with renal insufficiency who underwent mechanical mitral valve replacement in 2001 and transcatheter aortic valve replacement in 2021 due to severe aortic stenosis presented with chest pain and shortness of breath. Echocardiography suggested that the flow velocities of the mitral mechanical valve and aortic prosthetic valve were both within the normal range, with no significant paravalvular regurgitation; the tricuspid valve exhibited massive regurgitation (VMAX 258 cm/s, PGMAX 27 mmHg). Due to the high surgical risk, we simulated the procedure with a three-dimensional (3D)-printed model and performed transcatheter tricuspid valve replacement using a LuX-Valve (Ningbo Jenscare Biotechnology Co., Ningbo, China). DISCUSSION: We describe transcatheter tricuspid valve replacement using the LuX-Valve and preprocedural guidance with 3D printing. Postprocedural TR was significantly reduced, indicating that 3D printing plays an important role in preprocedural guidance and that the LuX-Valve was safe and practicable for tricuspid valve replacement.
No takes yet. Share an insight, caveat, or question.
Zhai et al. (2022) conducted a case report in Severe tricuspid regurgitation (n=1). Transcatheter tricuspid valve replacement using LuX-Valve with 3D printing guidance was evaluated on Postprocedural tricuspid regurgitation. Transcatheter tricuspid valve replacement using the LuX-Valve with 3D printing guidance successfully reduced postprocedural tricuspid regurgitation in a high-risk patient.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: