Key result
Evacuation of a trans-septal thrombus using the AngioVac aspiration system was successful, though complicated by right ventricle rupture requiring emergent surgical and ECMO support.
Why the study?
Does the AngioVac aspiration system under TEE and fluoroscopic guidance allow for the evacuation of trans-septal and pulmonary thrombi?
Case Report (n=1)
Does the AngioVac aspiration system under TEE and fluoroscopic guidance allow for the evacuation of trans-septal and pulmonary thrombi?
The AngioVac aspiration system under TEE and fluoroscopic guidance can be used for trans-septal clot evacuation, but carries significant risks of severe complications such as cardiac chamber rupture.
Answers are grounded in this paper and cite their sources.
May support transvenous AngioVac use for trans-septal thrombi; hypothesis-generating and requires prospective validation before adoption.
Surgical embolectomy is widely used to treat massive or submassive pulmonary thromboembolism and chronic thromboembolic pulmonary hypertension. [1]-[3] However, recently a new minimally invasive transvenous approach to clot evacuation using the AngioVac aspiration system (Vortex Medical, Norwell, Massachusetts) such as AngioVac aspiration system has been successfully used for retrieval of right atrial clot [4]-[6]. This is the first case report describing evacuation of trans-septal thrombus using an AngioVac aspiration system utilizing transesophageal echocardiograph (TEE) and fluoroscopic guidance. This case report also highlights the potential complications involved in AngioVac system which were readily diagnosed using TEE. We present a 66-year-old Caucasian female with a right atrial thrombus extending into left atrium through a patent for ovale (PFO) defect and a pulmonary artery thrombus. She underwent suction evacuation with an AngioVac aspiration system under TEE and fluoroscopic guidance. The right atrial thrombus extending into left atrium through a PFO was successfully evacuated through the PFO. However, the evacuation of the clot in the pulmonary artery was complicated by a rupture of the right ventricle requiring emergent sternotomy, cardiopulmonary bypass (CPB) and transient extracorporeal mechanical oxygenation (ECMO) support. The patient rapidly improved post-operatively and was discharged from hospital. Echocardiographic documentation of right heart thrombus and pulmonary emboli has poor prognostic implications. This report delineates the utility of TEE for visualization of cannula placement and real time aspiration of clots through the cannula. Potential complications associated with this technique are mechanical dislodge of thrombus from its attachment with subsequent embolization and rupture of cardiac chambers. Accessing the thrombus in the PA can be challenging and actual visualization of the cannula in the PA is often difficult and available only through TEE. This report highlights the utility of TEE for demonstrating trans-septal clot and the importance of precise positioning of cannula in avoiding complications. In spite of complications, AngioVac aspiration system under TEE and fluoroscopic guidance may become a safe alternative for surgical removal of intracardiac thrombi.
No takes yet. Share an insight, caveat, or question.
Ramamoorthy et al. (2017) conducted a case report in Right atrial thrombus extending into left atrium through a PFO and a pulmonary artery thrombus (n=1). AngioVac aspiration system with TEE and fluoroscopic guidance was evaluated on Evacuation of thrombus and complications. Evacuation of a trans-septal thrombus using the AngioVac aspiration system was successful, though complicated by right ventricle rupture requiring emergent surgical and ECMO support.
Synapse has enriched 10 related papers. Consider them for comparative context: